Ecol tests: FerroQ

  • • Standards: ASTM D7918-17a
  • • Unit: ppm
  • • Marking range: 0-200000
  • • Amount of grease: 3 g
  • • Application: plastic lubricants


  • method description:

    A sample in a Grease Thief or other sampler is introduced into the measuring chamber of the apparatus. The measurement is made by means of the distortion of the magnetic field generated by the coil of the apparatus. The tests are characterised by high accuracy and repeatability regardless of the distribution and size of the particles in the sample. The method is used for the testing of greases because of their consistency, where the particles, unlike in oils (liquids), do not move, and also because of the larger measuring range, as the content of iron particles in greases is usually much higher than in oils.



    interpretation of results:

    The FerroQ assay considers all ferromagnetic particles regardless of size. It is an excellent complement to the determination of iron in an emission spectroscopy study (RDE-OES where we determine particles below 5 micrometres in size. FerroQ correlation with iron tells us what fraction of iron is above 5 micrometres in size. A significantly higher ferroQ value in relation to iron can be indicative of the generation of large wear particles, not lubrication.

    check also

    You will find a list of all tests performed at the Ecol diagnostic laboratory on this page.

    contact